Only
Hayes provides your complete
soulution for high-quality repairs,
thorough cleaning, and
ongoing maintenance
Here's our 10-step
process for servicing your
high-speed handpiece:
3. DIAGNOSE THE PROBLEM
Test water spray, chuck strength and trubine
Test for air leaks
Check fiber optics
Look for any cracks or dents
Ream head if applicable
4.
DISASSEMBLE THE HANDPIECE IF APPLICABLE
1. SCHEDULE A FREE PICK UP
Free pick up and delivery
Discuss the problem you are having
2.
LOOK UP THE WARRANTY INFORMATION
Using our customized
warranty tracking
software, we
determine if the
handpiece is still
covered under our
repair warranty.
Remove all moving parts and components
5.
CLEAN THE HANDPIECE
INTERNAL CLEANING
Clean all components using proprietary
cleaning
techniques
Remove all debris including rust from the
drive air
tube, exhaust tube and chip air and chip water tubes
Clean internal head
EXTERNAL CLEANING
Clean head and remove dents when necessary
Clean and polish external shell and fiber optic ports
BEFORE
AFTER
6.
CHECK IMPELLER AND
SPINDLE FOR WEAR AND NICKS
7.TEST
FOR CONCENTRICITY
The
concentricity gauge
allows Hayes technicians
to prevent bur deflection
from exceeding .0011
inches, which reduces
stress and potential
damage to the
handpiece turbine.
Perform concentricity tests using propietary
equipment called a concentricity gauge
Maintain International Standards Organization
(ISO) standards of .0011 inches or less of
total indicator run-out (TIR)
Concentricity gauge
8. OVERHAUL HANDPIECE AND
REASSEMBLE
If the turbine is rebuildable,
we rebuild it with premium
parts and bearings to
O.E.M. standards
If the turbine failed
chuck strenght and /or
concentricity tests, we
install and replacement
turbine
Inspect and repair water
spray to ensure proper
atomization
Check water and chip air
tube
Replace O-rings
Replace fiber optics
if necessary
Replace end gasket
if applicable
Rethread if applicable
Replace the end
gasket if applicable
9.
TEST TORQUE, RPM and NOISE
10.
DELIVER THE HANDPIECE
AND REVIEW MAINTENANCE PROCDURES WITH YOUR STAFF
Discuss maintenance for high-speeds, low speeds,
swivels, scalers, fiber optics, and attachments
Review different types of lubricants
and cleaners
Check air pressure
for each handpiece
Review different
types of autoclaves
and related
procedures
Free maintenance demonstrations for staff
Hayes Handpiece of Hartford, Inc.
172 Bumstead Road Monson, Ma 01057 (800) 618-0110 FAX (413) 267-5080
Cell (413) 427-7438
Email: paulcote@verizon.net